The most compact ECP system available.

The features of the renew SPORT make it a major advancement in ECP systems

Renew™ SPORT Specifications

Treatment Pressure and Duration

  • Up to 6.0 psi
  • Treatment duration timer
    - Eleven pre-set treatment time options 10 - 60min

Renew™ SPORT Funcionality

  • Automatic firmware detection of the QRS complex and R-wave
  • Automatic ínflate / deflate start time setting ensures that counter pulsation occurs only during diastole. A push button switch allows manual fine-adjustment of the inflation / deflation times
  • Not filters and band-pass filters eliminate electrical interference to the cardiac signal
  • Software options provide ability to review treatment data or individual ECG events that occur during treatment or previous treatment sessions

Operating Environment

  • Temperature: 16ºC to 30ºC (60.BºF to 86ºF)
  • Humidity: 30% to 70% RH (non-condensing)
  • Barometric Pressure: B0kPa to 1 00kPa (equivalen! to 0m to 1,981 m elevation)
  • Absence of electromagnetic interference in the area (such as MRI device) Free of flammable gases and liquids, including anesthetics

Equipment Dimension and Weight

  • When ready for use: 1930mm (L) x 718mm (W) x 927mm (H)
  • When folded: 953mm (L) x 718mm (W) x 965mm (H)
  • Weight: 102kg (225lbs)

Power Requirement

  • 100V-240V at 50/60 Hz with a power rating of 1400W 3-wire electrical outlet depending on the country-specific requirements

Electical and Mechanical Safety Feature

  • Factory-set pressure relief valve to prevent patient exposure to excessive cuff pressures
  • Operator and patient access to an Emergency Stop button
  • Automatic deflation of cuffs when power loss of system shutdown occurs
  • Electrical isolation circuitry for patient protection
  • No activation of pump unless system software and laptop are operating

Software Safety Features

System will be alerted and inflation / deflation will be suspended in any of these events

  • lf heart rate exceeds the normal range of operation
    (< 36 BPM or> 124 BPM)
  • When an externa! (patient) ECG signal is not available or reliable
  • lf the laptop is unable to display data

Product images are for illustration purposes only. Actual product may vary.