renew™ SPORT Features
The most compact ECP system available.
The features of the renew™ SPORT make it a major advancement in ECP systems
Renew™ SPORT Specifications
Treatment Pressure and Duration
- Up to 6.0 psi
- Treatment duration timer
- Eleven pre-set treatment time options 10 - 60min
Renew™ SPORT Funcionality
- Automatic firmware detection of the QRS complex and R-wave
- Automatic ínflate / deflate start time setting ensures that counter pulsation occurs only during diastole. A push button switch allows manual fine-adjustment of the inflation / deflation times
- Not filters and band-pass filters eliminate electrical interference to the cardiac signal
- Software options provide ability to review treatment data or individual ECG events that occur during treatment or previous treatment sessions
Operating Environment
- Temperature: 16ºC to 30ºC (60.BºF to 86ºF)
- Humidity: 30% to 70% RH (non-condensing)
- Barometric Pressure: B0kPa to 1 00kPa (equivalen! to 0m to 1,981 m elevation)
- Absence of electromagnetic interference in the area (such as MRI device) Free of flammable gases and liquids, including anesthetics
Equipment Dimension and Weight
- When ready for use: 1930mm (L) x 718mm (W) x 927mm (H)
- When folded: 953mm (L) x 718mm (W) x 965mm (H)
- Weight: 102kg (225lbs)
Power Requirement
- 100V-240V at 50/60 Hz with a power rating of 1400W 3-wire electrical outlet depending on the country-specific requirements
Electical and Mechanical Safety Feature
- Factory-set pressure relief valve to prevent patient exposure to excessive cuff pressures
- Operator and patient access to an Emergency Stop button
- Automatic deflation of cuffs when power loss of system shutdown occurs
- Electrical isolation circuitry for patient protection
- No activation of pump unless system software and laptop are operating
Software Safety Features
System will be alerted and inflation / deflation will be suspended in any of these events
- lf heart rate exceeds the normal range of operation
(< 36 BPM or> 124 BPM) - When an externa! (patient) ECG signal is not available or reliable
- lf the laptop is unable to display data
Product images are for illustration purposes only. Actual product may vary.